Aerosol delivery from spacers in wheezy infants: a daily life study (2000) Open access

Pagina-navigatie:
Title Aerosol delivery from spacers in wheezy infants: a daily life study
Published in The European Respiratory Journal. ISSN 0903-1936.
Author Janssens, H.M. (Hettie); Heijnen, E.M.; Jong, de V.M.; Hop, W.C.J. (Wim); Holland, W.P. (Wim); Jongste, de J.C. (Johan); Tiddens, H.A.W.M. (Harm)
Date 2000-01-01
Language English
Type article
Abstract The aims of this study were to assess and compare dose delivery and dose variability of pressurized metered dose inhalers (pMDI)/spacers in wheezy infants in daily life and to investigate factors influencing aerosol delivery. In an open randomized crossover study in 25 wheezy infants aged 5-26 months, a metal spacer (Nebuchamber), a detergent coated (DC) and a non-detergent coated (nonDC) plastic spacer (Babyhaler) were tested at home for 7 days each. Budesonide (200 microg b.i.d) was administered via a Nebuchamber or fluticasone (125 microg b.i.d) via a Babyhaler. Aerosol was trapped in filters, positioned between the spacer and face mask. Cooperation was scored on diary cards. Electrostatic charge (ESC) of the spacers was measured. Evaluations of the administration technique were made from video recordings. Median (range) dose delivery of the filters expressed as per cent (%) of nominal dose, was 34% (3-59), 23% (1-49), and 41% (12-55) for the Nebuchamber, nonDC-Babyhaler, and DC-Babyhaler respectively. Considerable dose variability was found, median (range) within-subject dose variability, expressed as coefficient of variation, for the Nebuchamber (49% (15-249)) was significantly higher when compared with both nonDC- (36% (12-325)) and DC-Babyhalers (27% (10-122)), for which dose variabilities were similar. Detergent coating was effective to reduce electrostatic charge, and to increase dose delivery, but had no effect on dose variability. Bad cooperation was an important cause for high dose variability for all spacers (r=0.5-0.6, p<0.02). Many mistakes were made during the administration procedure.
Publication http://hdl.handle.net/1765/9563
Persistent Identifier urn:NBN:nl:ui:15-1765/9563
Metadata XML
Repository Erasmus University Rotterdam
Erasmus University Rotterdam

Go to page top
Go back to contents
Go back to site navigation